Bluelab Soil pH Pen Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Instrumentos de medida Bluelab Soil pH Pen. Bluelab Soil pH Pen User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 0
Care and use guide
Contents page
Features 2
Quick guide 2
Before  rst use begins 3
Removing and replacing the
storage cap/dibber 3
Hydration 4
IMPORTANT - Soil pH Pen probe care 4
To operate 6
Information about measuring the
pH of soils/media 7
Measuring soil pH value 9
Measuring pH value of a soil solution 10
Cleaning 11
Battery replacement 11
Calibration 12
Error messages 12
Troubleshooting guide 13
Technical speci cations 14
Bluelab pH probe care tools 14
Product guarantee 15
Contact details 15
a handy solution made easy
www.getbluelab.com
WEEE-Reg.-Nr. DE 33009570
Vista de página 0
1 2 3 4 5 6 ... 14 15

Resumo do Conteúdo

Página 1 - Care and use guide

Care and use guideContents pageFeatures 2Quick guide 2Before  rst use begins 3Removing and replacing thestorage cap/dibber 3Hydration 4IMPORTANT -

Página 2 - Quick guide

English10English108.0 Measuring the pH value of a soil solutionThe greatest source of error in soil analysis comes during sample collection. An eff

Página 3 - KCl Storage

English119.0 CleaningTo ensure accurate readings the probe tip needs to be kept clean at all times. Cleaning prior to calibration is necessary for

Página 4 - 3.0 Hydration

English1212.0 Error messagesThe following error messages appear for the following reasons.11.0 CalibrationpH calibration is required before fi rs

Página 5

English1313.0 Troubleshooting guideTrouble Reason CorrectionDrift - readingsslowly varyingGlassware not clean Clean glassware and calibrateWick cont

Página 6 - 5.0 To operate

English14Bluelab Probe Care - pHThe instrument is only as accurate as the probe is clean!Bluelab pH Probe KCl Storage SolutionThe perfect solution to

Página 7

English15If you need assistance or advice - we’re here to help you.Phone: +64 7 578 0849 Fax: +64 7 578 0847Email: [email protected]

Página 8

English2FeaturesDirect pH and temperature measurements from soils/mediaSelectable units for temperatureAutomatic Temperature Compensation (ATC) Backli

Página 9 - Ensure you rinse the probe

English31.0 Before  rst use beginsATTENTION: Salt-crust forming on the pH probe is normal. A small amount of KCl solution from the storage cap/dib

Página 10

English44.0 IMPORTANT - Soil pH Pen probe careTo ensure you receive a long life from your Soil pH Pen, please ensure you follow the guide below. Th

Página 11 - 10.0 Battery replacement

English54.0 IMPORTANT - Soil pH Pen probe care cont.The soil pH probe does not last forever• Probes do not last forever. They age through normal us

Página 12 - 11.0 Calibration

English65.0 To operateTurn Soil pH Pen onPress power button. The last measurement is recalled for 3 seconds.To turn Soil pH Pen offPress and hold t

Página 13 - 13.0 Troubleshooting guide

English76.0 Information about measuring the pH of soils/mediapH is the measurement of the hydrogen ion concentration (H+) - acidity and its opposit

Página 14 - Solution

English86.0 Information about measuring the pH of soils/media cont.Taking pH measurements of soils with an electronic meter is indicative rather th

Página 15 - Limitation of Liability

English97.0 Measuring soil pH valueFollow the steps below to take pH measurements of soil and media.Remove the top layer of soil/media from the sur